Electric Techniques
Electric Force Microscopy
282 Dimension 3100 Manual Rev. D
Analog 2 Voltage Applied to the Sample (No Extender Electronics)
The jumper configuration in Figure 16.2g connects the Analog 2 signal from the NanoScope III
controller (± 12 VDC range) to the sample chuck. Remember to enable the Analog 2 voltage line as
described in Section 16.2.2.
Figure 16.2g Jumper Configuration (Application of Analog 2 Voltage to Sample)
Sample Chuck Ground/Bias
Analog 2
Gain Select
Auxiliary D (to NanoScope III controller)
Ground
STM Signal (from Dimension head)
Indicates jumpers
Unused
Ground
Ground
AFM Tip
Sample Chuck
Sample
Tip
Analog 2